SD003-PCR-LA (Quick Immunity Sequencer) – Updater

SD003-PCR-LA

SD003-PCR-LA (Quick Immunity Sequencer) Updater

This software is to update the application version of the SD003-PCR-LA for power line disturbance immunity testing.

Ver.

1.13.00

Last update

2015/7/23

Configuration

  • Vista(x86/x64)
  • 7(x86/x64)
  • 8(x86/x64)
  • GPIB RS232C

Download

Release Note

VER 1.13
2015年7月16日

IEC61000-4-29

Resolved issues
There was a problem in which sometimes High-impedance / Low-impedance of "DC short interruption" was not output as configured.
Resolved issues
When "Run" was carried out several times in output ON, there was a problem in which interruption of setting become rarely longer than its configuration.
VER 1.12
2013年12月16日

IEC61000-4-29

Resolved issues
There was a problem in which interval time was longer for 0.001 second than the setup time when repeating.
VER 1.11
2013年4月2日

IEC61000-4-28

Resolved issues
At f1+Delta f, there was a problem in which minus value was out of range. This problem occurred only by Ver1.10.
VER 1.10
2013年5月25日

IEC61000-4-28

Resolved issues
There was a problem in which a command error occurred when the value of more than 999.999 seconds was set.
VER 1.09
2012年2月17日

Quick Immunity Sequencer

Resolved issues
There was a problem that runtime was insufficient in the installer. This problem occurred only by Ver1.08.
VER 1.08
2011年1月1日

IEC61000-4-13

Resolved issues
We corrected a problem of effective value of the output voltage being different which was caused by the fact that we were not converting the voltage distortion factor due to the superimposed harmonics by Over swing, Sweep in frequency, Odd non multiple of 3, Odd multiple of 3 and Even harmonics.
VER 1.07
2009年11月30日

IEC61000-4-13

Resolved issues
At Flat curve, there was a problem that Ver1.0.6 generated overshoots and undershoots when switching between banks.
VER 1.06
2009年11月12日

IEC61000-4-13

Resolved issues
There was a problem that Level 0.80 required by Flat Curve Class 3, could not be set under the three-phase system.
Resolved issues
At Flat Curve under the three-phase system, the output waveform has been corrected from "line-to-line" peak-clip to "inter-phase" peak-clip.
VER 1.05
2009年4月8日

Quick Immunity Sequencer

Resolved issues
When using in European region where "comma" is used as the decimal point, there was a problem that the error message "x.xx is not a valid float" was not correctly shown.
VER 1.04
2007年6月4日

Quick Immunity Sequencer

Resolved issues
There was a problem that unwanted phase-shifting was generated when using IT01.
Resolved issues
There was a problem that unexpected "End" could be detected before sequence start after operating Send → Run.
Resolved issues
When launching at the condition of "Not use IT01", changing between "Use IT01" and "Not use IT01" caused a communication error.
Resolved issues
There was a problem that communication errors caused unwanted application's abortion.
VER 1.02
2005年5月13日

Quick Immunity Sequencer

Resolved issues
There was a problem that send operation caused hang-up after switching "Use IT01" to "Not use IT01".
VER 1.00
2004年4月7日

Quick Immunity Sequencer

New Feature
The first release