Update Module for Quick Immunity Sequencer 2

This is the latest update module for Quick Immunity Sequencer 2. To install this module, your PC must already install Quick Immunity Sequencer 2.

Ver.

4.1.0

Last update

2021/6/9

Configuration

  • 10 (x86/x64)

Notes

*Attention when upgrading to SD009-PCR-LE_Ver337 or later.

Download

Release Note

Ver4.1.0
April 08, 2021

Quick Immunity Sequencer 2 / Wave Bank Memory

New Feature
This version operates also in AC power supply PCR-WEA series.
Resolved issues
When the Setting Limits of Instruments menu was opened, there was a problem of which an error occurs in Setting Limits dialog.
Improvement
If AC power supply detect the command error of over-range etc. during sending program data for test start, test will be stopped.
Improvement
Even if it is in simulation mode, reading and drawing are able to perform from waveform data file.
Ver 4.0.4
December 24, 2020

Quick Immunity Sequencer 2

Improvement
Operation manual was updated.
Ver 4.0.3
December 25, 2019

Quick Immunity Sequencer 2 / Wave Bank Memory

Resolved issues
There was a problem which did not match progress with an actual progress bar only at the time of DSI use.
Resolved issues
There was a problem on which the function to return sensing mode and a response at the time of PCR-WE use has not been operated.
Updated
There was a change which lowers the minimum of voltage to 5V from 50V.
Updated
When the error of cable connection, FUSE, and OHP occurred in DSI, there was a change which stop the test and turn off the output of PCR-LE/WE.
Resolved issues
There was a problem on which the waveform of Ua-Ub differed from the preview waveform by 3 phase dip.
Resolved issues
Correction has been made that only numerical characters could be input.
Ver 4.0.2
September 18, 2019

Quick Immunity Sequencer 2

Resolved issues
There was a problem out of which an error occurs at the time of first time starting of this application.
Ver 4.0.1
June 27, 2019

Quick Immunity Sequencer 2

Improvement
This version comply with "IEC61000-4-13 Meister curve 2015".
Ver 4.0.0
June 03, 2019

Quick Immunity Sequencer 2

New Feature
This version operates also in PCR-WE series.
Ver 3.3.7
February 20, 2018

Quick Immunity Sequencer 2

Resolved issues
The decimal point (comma) of Europe is adapted.
Ver 3.3.6
March 28, 2016

Quick Immunity Sequencer 2

New Feature
The response options was added to the options menu. Setup for response of PCR-LE in IEC61000-4-29 and other standards were added. Moreover, added a function that sensing mode and response were returned to last setting when application exit.
Ver 3.3.5
March 03, 2016

Quick Immunity Sequencer 2

Resolved issues
IEC61000-4-29 There was a problem which does not become High Impedance by the DC Short interruptions. As this solution serge suppression function turned OFF. Firmware Ver4.50 of PCR-LE becomes effective to this function.
Ver 3.3.4
August 28, 2015

Quick Immunity Sequencer 2 Wave Bank Memory

Improvement
This version operates in Windows 10.
Ver 3.3.3
April 08, 2015

Quick Immunity Sequencer 2

Improvement
IEC61000-4-17 The setting lower limit of DC voltage (Udc) was changed into 10.0V from 50.0V.
Ver 3.3.2
October 20, 2014

Quick Immunity Sequencer 2

Resolved issues
There was a problem in which communication error was occurred, when DSI series was connected by RS232C.
Ver 3.3.1
May 23, 2014

Quick Immunity Sequencer 2

Resolved issues
IEC61000-4-13 There was a problem in which command error occurred during a test of the meister curve. (Only LAN hislip)
Ver 3.3.0
April 07, 2014

Quick Immunity Sequencer 2

New Feature
This version operates also in DSI series.
Ver 3.2.0
March 22, 2013

Quick Immunity Sequencer 2

New Feature
The sending progress bar was added.
New Feature
The visual progress of test was added.
New Feature
The following item was added to the option menu. Show confirmation Message of change setting - Change Setting that affect soon the output value - Change Seq Tab value Display test progress bar Test result file is not saved
Ver 3.1.1
February 18, 2013

Quick Immunity Sequencer 2

Resolved issues
IEC61000-4-11, IEC61000-4-34 There was a problem in which a command error occurred at Phase voltage on the Voltage dips.
Ver 3.1.0
February 01, 2013

Quick Immunity Sequencer 2

Resolved issues
IEC61000-4-11, IEC61000-4-34 There was a problem in which the angle of Ua phase did not shift at Line voltage in three-phase.
Resolved issues
IEC61000-4-11, IEC61000-4-34 There was a problem in which the angle of each phase did not ramp in the part of the fall and rise of the Voltage variation at Line voltage in three-phase.
Ver 3.0.2
December 20, 2012

Quick Immunity Sequencer 2

Improvement
IEC61000-4-27 Setup of the reference phase is added and enabled the reference phase to change.
Improvement
IEC61000-4-11, IEC61000-4-34 In single-phase 3-wire or if Phase select is set to Line voltage in three-phase, the phase voltage that you enter is converted to the nearest line voltage and displayed under the phase voltage.
Improvement
IEC61000-4-11, IEC61000-4-34 The cycle value of Dips and Short that you enter is converted to the nearest sec value and displayed under the cycle value.
Ver 3.0.1
November 27, 2012

Quick Immunity Sequencer 2

Improvement
IEC61000-4-11, IEC61000-4-34 When the Line voltage is selected, the waveform preview can be drawn on the line voltage.
Improvement
IEC61000-4-14 When the first Seq is selected at Interval, the phase angle of start is 0 degree.
Improvement
IEC61000-4-28 The phase angle of start in the first Seq is 0 degree.
Resolved issues
There was a problem in which Setting Limits did not correspond to the parallel system.
Ver 3.0.0
October 23, 2012

Quick Immunity Sequencer 2

New Feature
This version operates also in single phase - three wire system.
Improvement
Improved to make zooms in a waveform preview one more step.
Improvement
Improved to make a waveform preview similar to an output waveform.
Resolved issues
There was a problem which had disorder in the waveform preview depend on a setting value.
Resolved issues
There was a problem which the waveform preview did not correspond to the voltage between lines.
Ver 2.0.1
October 09, 2012

Quick Immunity Sequencer 2

Resolved issues
In the case of IEC61000-4-34, there was a problem that the error message "Setting voltage is out of range" was displayed even if setting voltage was within the limits.
Resolved issues
There was a problem in which IEC61000-4-11 Short interruption did not operate when IT01-PCR-L was used.
Ver 2.0.0
June 20, 2012

Quick Immunity Sequencer 2 / Wave Bank Memory

New Feature
This version operates also in three phase system.
New Feature
This version operates also in three phase system.
Ver 1.0.1
February 17, 2012

Quick Immunity Sequencer 2 / Wave Bank Memory

Improvement
English contents and user's manual can be now referenced from the Help menu.
Improvement
English contents can be now referenced from the Help menu.
Ver 1.0.0
December 26, 2011

Quick Immunity Sequencer 2 / Wave Bank Memory

New Feature
The first release.
未設定メニュー

Update Module for SD012-PCR-LE/WE (Avionics Test Software)

This is the latest update module for SD012-PCR-LE/WE (Avionics Test Software). To install this module, your PC must already install SD012-PCR-LE/WE (Avionics Test Software).

SD012-PCR-LE (Avionics Test Software)

Ver.

1.1.0

Last update

2021/2/13

Configuration

  • Applied Products: SD012-PCR-LE Ver.1.X
  • Windows 7 (x86/x64)
  • Windows 8 (x86/x64)
  • Windows 10 (x86/x64)

your PC must already install SD012-PCR-LE (Avionics Test Software).

Download

Release Note

Ver 1.1.0
November 17, 2020

Avionics Test Software

New Feature
Added the display of the currently set "output mode" to the main screen.
New Feature
Added "Show message box" to the options.
Resolved issues
Fixed a bug that the previous test conditions were not reflected when the application was restarted immediately after loading the User file.
Ver 1.0.5
March 12, 2020

Avionics Test Software

Resolved issues
Added in DO-160F,G 16.5.1.2 because the number of test levels was small.
Resolved issues
In DO-160F,G 16.5.1.3, the test step frequency was half of the standard value, so there was an error in the number of corrections and repetitions.
Resolved issues
There was an error in the repetition of DO-160F, G standard test data.
Resolved issues
In DO-160F,G 16.5.2.3.2, there was an error in the interval time.
Resolved issues
In DO-160F,G 16.5.1.8, the arbitrary waveforms of A (WF) and A (CF) / A (NF) were the same.
Updated
In DO-160G 16.6.1.3b, Table16-3 has changed.
Resolved issues
In DO-160F,G 16.6.1.3c, 16.6.1.3d, 16.5.1.4b, c, d, the interval time was corrected to 10 seconds.
Resolved issues
In DO-160F,G 16.6.1.3d, the setting voltage range was incorrect.
Resolved issues
In DO-160G 16.5.1.4, 16.5.1.4b, 16.5.1.4c, the setting voltage range was incorrect.
Resolved issues
In DO-160F 16.5.1.4c, the setting voltage range was incorrect.
Ver 1.0.4
March 12, 2019

Avionics Test Software

New Feature
In this version, test conditions were subdivided for each TestNo.
Ver 1.0.3
April 21, 2015

Avionics Test Software

Resolved issues
List of Errors pages were added to Operation Guide.
Ver 1.0.2
July 01, 2014

Avionics Test Software

New Feature
In this version, I added function for being able to select all listed tests.
Resolved issues
A Voltage range setting of MIL-STD-704 LDC test data had a problem.
Ver 1.0.1
January 09, 2014

Avionics Test Software

Resolved issues
A Loop setting of MIL-STD-704E and F test data had a problem.
Resolved issues
An unnecessary test was included in MIL-STD-704E.
Resolved issues
A Loop setting of DO-160F and G test data had a problem.
Resolved issues
A Loop setting of JIS W0812 test data had a problem.
Ver 1.0.0
August 07, 2013

Avionics Test Software

New Feature
The first release.

SD012-PCR-WE (Avionics Test Software)

Ver.

1.1.0

Last update

2021/2/13

Configuration

  • Applied Products: SD012-PCR-WE Ver.1.X
  • Windows 7 (x86/x64)
  • Windows 8 (x86/x64)
  • Windows 10 (x86/x64)

your PC must already install SD012-PCR-WE (Avionics Test Software).

Download

Release Note

Ver 1.1.0
December 17, 2020

Avionics Test Software

New Feature
Added the function of "Setting the Wiring Method of the AC Power Supply".
New Feature
Added the display of the currently set "output mode" to the main screen.
New Feature
Added "Show message box" to the options.
Resolved issues
Fixed a bug that the previous test conditions were not reflected when the application was restarted immediately after loading the User file.
Ver 1.0.0
March 12, 2020

Avionics Test Software

New Feature
The first release.
未設定メニュー

Firmware Updater

This software is to update the firmware version of the PCR-LE2 Series.

Ver.

5.07

Last update

2020/11/4

Configuration

  • Applied Products : PCR-LE2 Series – All models
  • Applied Firmware Version : ver4.00 or later

Caution: If you update a PCR6000LE2 or PCR9000LE2 whose firmware version is 4.00, the DC output in three-phase AC+DC mode will not meet the specifications. If you want to use three-phase AC+DC mode after updating the firmware, the PCR-LE2 hardware must be adjusted. For this adjustment, contact your Kikusui agent or distributor.

Notes

  • Confirm the firmware version of the PCR-LE2.
    *You can confirm the firmware version by the CONFIG (SHIFT + OPR MODE) > MODEL ID (F5).
  • If any data is stored in the internal memory, save the data to the USB memory before updating the firmware version.
  • The USB memory (up to 16GB) that recognizes the PCR-LE2 is required for the firmware update
    *To confirm the applicable USB memory : Insert the USB memory to the USB port of the PCR-LE2, then check for the display. If the “USB Memory Found” shows on the display, the USB memory can be used for the firmware update. If the USB can not be recognized by the PCR-LE2, nothing will be displayed.
  • In case of failure for the installation of firmware version update, please contact your local distributor.

Download

Related Files

Please download the files below.

Firmware update guide

Last update

2014/2/3

Configuration

Adobe Reader

Download

CD image (Operation manual)

Last update

2017/10/19

Download

Firmware Updater Release Note

VER 5.07
September 09, 2020

PCR-LE firmware

Improvement
The protection function when the output voltage range is L range and Vcc is FIXed has been improved.
VER 5.06
March 25, 2020

PCR-LE firmware

Resolved issues
When a protection function was activated by the detection of outage, processing to invalidate the output control from SCPI command has been added.
VER 5.05
January 28, 2020

PCR-LE firmware

Resolved issues
When the SD021-PCR-LE reconnects communication with the PCR-LE, the problem that setting frequency became the input frequency (50Hz or 60Hz) has been corrected.
VER 5.03
November 22, 2018

PCR-LE firmware

Resolved issues
When the SD021-PCR-LE reconnects communication with the PCR-LE, the problem that setting frequency became the input frequency (50Hz or 60Hz) has been corrected.
VER 5.02
October 10, 2017

PCR-LE firmware

Resolved issues
The sudden phase change feature of the sequence function has been corrected.
VER 5.01
September 08, 2017

PCR-LE firmware

New Feature
The single-phase three-wire/three-phase system (option) feature has been improved. The master-slave parallel operation (option) has been improved. Firmware version 5.xx is not compatible with version 4.xx or earlier. As such, a firmware version check function that runs when a multi-phase operation system or parallel operation system is turned on has been added. If the PCR-LE system contains a mixture of version 5.xx and version 4.xx or earlier versions, “TRBL-22 VERSION ERROR” will occur when the power is turned on.
New Feature
An aperture time setting function has been added. * In conjunction with this addition, the following functions have been changed or deleted. - The measurement update interval of measured values has been changed to aperture time (except line voltage measurements and harmonic current analysis data). - The measurement averaging function has been deleted. - The estimated minimum frequency function has been deleted (a function that was available when an analog signal interface option (EX05-PCR-LE) was installed).
Updated
The soft sensing correction range has been changed from +10% to ±10%.
Updated
Frequencies less than 40 Hz have been excluded from the Soft sensing operation correction range.
Improvement
A function that displays the soft sensing control status on the front panel has been added.
Improvement
The protection function processing that takes place when a sensing wire connection error is detected has been changed. When the protection function is activated, “ALM-22 SENSING FAILURE” will occur. To clear the sensing wire connection error, use alarm clear. * However, depending on the hardware version, the previous error (TRBL-22 SENSING FUSH TRIP) may occur. In this case, the method to clear the sensing wire connection error is the same as before (turn off the power switch and then back on).
VER 4.53
May 22, 2017

PCR-LE firmware

Resolved issues
When a sequence is ended in AC+DC mode, the problem of the voltage not being set to the AC voltage specified in the last step due to the peak voltage limit when the preset DC voltage before starting the sequence + the AC voltage waveform’s peak value set in the last step falls outside the -431 V to 431 V range (H range) or -215.5 V to 215.5 V range (L range) has been fixed.
VER 4.52
August 10, 2016

PCR-LE firmware

Resolved issues
The problem in which a portion of the output waveform was occasionally different from the voltage settings when a switch was made to a waveform bank containing waveform data with different phases or amplitudes has been fixed.
Resolved issues
The single-phase three-wire/three-phase system (option) feature has been corrected. The problem in which the PCR-LE was initialized to factory default settings when the power supplies of all phases were restarted after only the U-phase unit was turned on and a communication error (TRBL-24:U-V COMM MISS) occurred between the U phase and V phase has been corrected.
VER 4.51
June 17, 2016

PCR-LE firmware

Resolved issues
The problem in which a portion of the output voltage waveform was occasionally different from the voltage setting when the frequency setting was changed has been fixed.
Resolved issues
The problem in which the generated peak clipped waveform data being initialized and set to a sine wave when the power switch was turned off and then back on has been fixed.
Resolved issues
The single-phase three-wire/three-phase system (option) feature has been corrected. The problem in which only the V-phase or W-phase waveform data of the arbitrary waveform data set with the SD011-PCR-LE being initialized and set to a sine wave when the power switch was turned off and then back on has been fixed.
VER 4.50
January 15, 2016

PCR-LE firmware

New Feature
A voltage surge suppression function setting has been added. A setting for turning off the voltage surge suppression function when the output is turned off has been added.
VER 4.43
November 25, 2015

PCR-LE firmware

Improvement
The output-off response during Vcc wait times when the internal Vcc voltage is set to auto has been improved.
Resolved issues
The problem of the sequence not changing to the frequency of the next step when frequency "RAMP ON" is specified with the sequence function and the frequencies of the preceding step and the following step are the same has been fixed.
Resolved issues
The single-phase three-wire/three-phase system (option) feature has been corrected. The problem of not being able to set the second parameter (ON/OFF) of PROG:EDIT:PHAS:UV and PROG:EDIT:PHAS:UW (results in a parameter error when this is attempted) has been fixed.
VER 4.42
June 23, 2015

PCR-LE firmware

Resolved issues
A problem with the PCR-LE2 500Hz LMT model has been fixed. The problem that remote control settings were initialized when the instrument was power-cycled in order to switch from single-phase-three-wire output to three-phase output,has been fixed.
VER 4.41
May 26, 2015

PCR-LE firmware

New Feature
The PCR-LE now returns the installed analog signal interface ("EX05" or "EX06") in response to the communication commands SYST:OPT? and *OPT?.
Resolved issues
Correction to a problem that occurred in master-slave parallel operation (option) and during PCR-LE2 single-phase output. The problem of the QUEStionable status register being cleared immediately when an overcurrent protection was activated has been fixed.
VER 4.40
April 10, 2015

PCR-LE firmware

Improvement
The overload protection (ALM-06: OVERLOAD) detection has been improved to accommodate repeated overload conditions. A full scale bar, which indicates that the internal temperature is high, has been added to the load meter. The time (busy state) during which the output cannot be turned on after an ALM-03:OCP occurs has been changed from 10 seconds to 120 seconds. The time (Busy state) during which the output cannot be turned on after an ALM-06:OVERLOAD occurs has been changed from 30 seconds to 120 seconds.
Improvement
An indication of the time until the busy state is cleared (until the output can be turned on) has been added.
Improvement
An indicator that appears when the internal semiconductor protection is activated has been added to the front panel.
Improvement
The maximum used output voltage range (the lower limit) when a fixed internal Vcc is used has been changed. A tracking process in which the maximum used output voltage is kept higher than the peak output voltage has been added.
Improvement
A power unit failure diagnosis at power-on has been added.
Improvement
In AC mode when the voltage settings is at or lower than 10 V (L range) or 20 V (H range), the time that must elapse before the output is turned off when the current limit is exceeded has been changed.
Resolved issues
The single-phase three-wire system (option) has been corrected. Fixed a problem that was preventing panel setting files saved to USB memory devices from being recalled.
Resolved issues
The parameter MAX value for the SYST:CONF:PHAS:UV and SYST:CONF:PHAS:UW communication interface commands has been corrected to 359.
Resolved issues
Fixed a problem that was not resetting the synchronization function to OFF during factory default initialization.
Resolved issues
Fixed the time until ALM-03:OCP is generated after the internal semiconductor protection is activated in DC mode or AC+DC mode.
Resolved issues
Fixed a problem in the peak voltage range check process that is executed when memory is recalled in AC+DC mode.
VER 4.36
November 28, 2014

PCR-LE firmware

Improvement
The internal semiconductor protection (OCP) detection has been improved.
Improvement
The reliability in the internal communication of the parallel operation driver (PD05M-PCR-LE/PD05S-PCR-LE) has been improved.
VER 4.34
October 03, 2014

PCR-LE firmware

Improvement
A filtering function has been added for the detection of alarm signals.
Resolved issues
A problem that had been occurring when an analog signal interface option (EX05-PCR-LE) was installed has been corrected. A problem with communication interface commands that had been rejected due to "-221 Setting Conflict" errors has been corrected. FUNC:SOUR / SYST:CONF:EXT:SINP:POL / SYST:CONF:EXT:EST:FREQ:LOW
VER 4.33
April 14, 2014

PCR-LE firmware

Resolved issues
The output waveform bank could not be changed when the output method was single-phase output. This has been corrected.
Resolved issues
In VER 4.20 to VER4.32, PCR-LE/LE2 memory files saved to USB memory could not be recalled. This has been corrected.
VER 4.32
March 03, 2014

PCR-LE firmware

Resolved issues
PCR-LE2 corrections / The selected output method was indicated as single-phase when the output method was changed from single-phase three-wire to three-phase and then the PCR-LE2 was restarted. This has been corrected.
Resolved issues
When phase voltage was specified in single-phase three-wire DC mode, conversion to line voltage could not be carried out. This has been corrected.
Resolved issues
If the output was turned on immediately after a sequence was stopped with the OUTPUT key, the PCR-LE/LE2 was generating the voltage specified before the sequence execution. This has been corrected so that PCR-LE/LE2 generates the voltage at the end of the sequence.
VER 4.31
January 24, 2014

PCR-LE firmware

Resolved issues
Depending on the settings, overload protection had been activating even at output less than the rated current. This has been corrected.
Resolved issues
The first soft start after power-on had not been working. This has been corrected.
VER 4.30
November 28, 2013

PCR-LE firmware

New Feature
Support for the SD019-PCR-LE/SD020-PCR-LE
Resolved issues
The three-phase system (option) feature has been corrected. In AC+DC mode, W phase could not be selected in the phase selection screen shown on the front panel. This has been corrected.
Resolved issues
A problem that had been occurring when an analog signal interface option (EX05-PCR-LE) was installed has been corrected. The signal source, polarity, and estimated minimum frequency settings had not been backed up. This has been corrected.
VER 4.20
September 19, 2013

PCR-LE firmware

New Feature
A single-phase three-wire/three-phase system (option) feature has been added. AC+DC mode for multi-phase output
New Feature
Analog signal interface option (EX05-PCR-LE/EX06-PCR-LE) has been added. SHUTDOWN has been added to bit 5 of the QUEStionable status register.
Resolved issues
UVP and OVP judgment in AC+DC mode has been changed to RMS judgment. In conjunction with the change in judgment to RMS values, the OVP and UVP setting ranges have been changed. AC mode / AC+DC mode: 0 to 474.1 V
Resolved issues
The single-phase three-wire/three-phase system (option) feature has been corrected. When trigger input standby and starting phase angle were set in the same step of a sequence, the output waveform of the V phase (W phase) was shifted one step earlier. This has been corrected. If the internal VCC voltage was fixed, there were cases where the output voltages could not be set at once. This has been corrected.
Resolved issues
A problem that occurred when an analog signal interface option (EX05-PCR-LE) was installed in a single-phase three-wire/three-phase system (option) has been corrected. Signal source status display, EXT.OPT menu operation, and EXT.SIG menu display have been corrected.
Resolved issues
Overload was not activated even when the rated current was exceeded when reverse current was in use. This issue has been corrected.
Resolved issues
When a sequence was executed with the internal VCC voltage set to auto, there were cases where the VCC voltage did not change automatically and the output waveform became distorted. This has been corrected.
VER 4.00
April 01, 2013

PCR-LE firmware

New Feature
A single-phase three-wire/three-phase system (option) feature has been added. Resistance display of each phase that corresponds to the output impedance setting that ranges from 1 to 100% has been added. U phase offset and phase difference ramp settings have been added to the sequence feature. A U phase offset cancel feature has been added.
VER 3.20
February 01, 2013

PCR-LE firmware

Resolved issues
The expression for calculating the measured power factor has been changed from a signed expression to an unsigned expression. (λ=P/S → λ=|P|/S)
Resolved issues
When the power is turned off, the panel’s OUTPUT OFF indication and the output status sometimes did not match. This has been corrected.
VER 3.12
December 10, 2012

PCR-LE firmware

Improvement
The reliability in the internal communication of the parallel operation driver (PD05M-PCR-LE/PD05S-PCR-LE) has been improved.
Resolved issues
The load meter full scale calculation when the current limit is set close to the minimum value has been corrected.
VER 3.10
November 12, 2012

PCR-LE firmware

New Feature
Support for the parallel operation driver option / PD05M-PCR-LE (master) / PD05S-PCR-LE (slave)
New Feature
Support for the analog signal interface option / EX05-PCR-LE / EX06-PCR-LE
New Feature
The 2P Mode On/Off function (for selecting the output mode) has been added to the AC mode that uses the single-phase, three-wire output driver option (2P05-PCR-LE).
New Feature
A communication interface command for querying the measured AC power (overall reactive power) has been added. (for the single-phase, three-wire output and three-phase output options) FETCh: POWer:AC:REACtive:TOTal$3 / MEASure:POWer:AC:REACtive:TOTal
Improvement
When panel operation is not possible due to the current condition or conflicts with other settings, the reason is now indicated for better operability.
Resolved issues
The reversed UVP/OVP judgment polarity on the V phase side when the single-phase, three-wire output driver option (2P05-PCR-LE) is in use has been fixed.
Resolved issues
The overload judgment point (threshold) when the frequency is set less than 40 Hz in AC mode has been corrected.
Resolved issues
The error in the calculation of the reactive power of the V and W phases when the single-phase, three-wire output or three-phase output option is in use has been fixed.

未設定メニュー

Update Module for Wavy for PCR-LE

This is the latest update module for Wavy for PCR-LE. To install this module, your PC must already install Wavy for PCR-LE Ver.5.4.X.

Ver.

5.6.0

Last update

2019/3/22

Configuration

  • 7 (x86/x64)
  • 10 (x86/x64)

your PC must already install Wavy for PCR-LE Ver.5.4.X.

Download

Release Note

Ver 5.6.0
March 22, 2019

SD011-PCR-LE

New Feature
In this version, the function was added so that AC+DC mode can be selected at 1P3W and 3P.
Ver 5.5.3
April 22, 2015

SD011-PCR-LE

Resolved issues
List of Errors pages were edited to Operation Guide.
Ver 5.5.2
March 09, 2015

SD011-PCR-LE

Resolved issues
The detection of the alarm (internal semiconductor protection) had a problem.
Ver 5.5.1
June 18, 2014

SD011-PCR-LE

Resolved issues
An end state judgment of the sequence had a problem.
Ver 5.5.0
May 19, 2014

SD011-PCR-LE

New Feature
In this version, I added phase sweep setting to a sequence.
New Feature
In this version, I added a sequence preview screen.
Ver 5.4.1
November 12, 2012

SD011-PCR-LE

Resolved issues
Remote control panel window line voltage conversion indication had a problem. (single-phase 3 wire)
Resolved issues
The set point reflection from the PCR-LE at the time of the remote control panel window start had a problem.
Resolved issues
The number of the input letters of the interface window IP address had a problem.
Ver 5.4.0
September 10, 2012

SD011-PCR-LE

New Feature
This version operates also in single-phase 3 wire and three phase system.
Ver 5.3.0
July 02, 2012

SD011-PCR-LE

New Feature
The first release.
未設定メニュー

アップデート

本ソフトウェアは、SD019-PCR-LE/SD020-PCR-LE(Wavy Smart Grid Edition) のためのアップデートファイルです。

バージョン

1.2.1

更新日

2018/08/06

対象製品

  • SD019-PCR-LE Ver.1.0/SD020-PCR-LE Ver.1.0

環境

  • 7 (x86/x64)
  • 8 (x86/x64)
  • 10 (x86/x64)

SD019-PCR-LE Ver.1.0/SD020-PCR-LE Ver.1.0 が、PCに既にインストールされている必要があります。
インストール後、Wavyのバージョン情報を確認すると、Ver.1.2.1となります。

ダウンロード

    リリースノート

    Ver 1.2.1
    August 03, 2018

    Grid Simurator

    Resolved issues
    単相2線時に直接制御画面が起動できない問題があった。
    Ver 1.2.0
    March 18, 2016

    Grid Simurator

    New Feature
    2015年10月改定試験方法対応
    Ver 1.1.2
    March 09, 2015

    Grid Simurator

    Resolved issues
    PCR-LE本体の内部半導体保護アラームを検出できない問題があった。
    Ver 1.1.0
    October 23, 2014

    Grid Simurator

    New Feature
    2014年7月改定試験方法対応(複数直流入力システム対応)
    New Feature
    シーケンス作成支援機能追加(三相不平衡電圧)
    Ver 1.0.3
    August 22, 2014

    Grid Simurator

    Resolved issues
    三相時の位相変化を伴う瞬時電圧低下試験(FRT試験)方法を位相急変から二相短絡に修正した。
    Ver 1.0.2
    June 18, 2014

    Grid Simurator

    Resolved issues
    モニタログファイルを保存しない場合、シーケンス終了時に実行状態が終了にならない問題があった。
    Ver 1.0.1
    May 21, 2014

    Grid Simurator

    New Feature
    2013年11月改定規格対応(JETGR0003-1-3.1(2013)太陽光発電システム用)
    New Feature
    2013年11月改定規格対応(JETGR0003-4-2.1(2013)多数台連系対応型太陽光発電システム用)
    Ver 1.0.0
    October 31, 2013

    Grid Simurator

    New Feature
    初版リリース
    未設定メニュー

    IVI-COM Multi-Environment Drivers

    An IVI-COM multi-environment instrument driver compliant with the IVI specifications. Compatible with Microsoft Office VBA, Visual Basic 6.0, Visual Basic.NET. And because an IVI-C driver is bundled, it is possible to use it easily in LabVIEW and LabWindows/CVI.

    Ver.

    1.1.0.0

    Last update

    2018/6/29

    Configuration

    • 7(x86/x64)
    • 8(x86/x64)
    • 10(x86/x64)
    • LAN/USB/GPIB/RS-232C
    • x64:For 64bit OS, PCR-LE2 or PCR-LE Three-Phase operation.
    • x86:For 32bit OS, PCR-LE2 or PCR-LE Three-Phase operation.

    Notes

    • If KikusuiPcrle 3_1_0_0_0(x64/x86) has been installed, uninstall it and install this driver (1.1.0.0).
    • This driver requires the IVI Shared Component since version 2.3.
    • To use IVI-COM instrument drivers, a VISA Library that supports VISA COM.
    • We recommend using one of the VISA Libraries supporting VISA COM shown below. (Mind that multiple versions of VISA cannot be installed at a time.)
      • NI-VISA 5.1.0 or later
      • Agilent IO Libraries Suite 16.0 or later
      • KI-VISA 5.0.5 or later
    • When using 32bit OS, the x86 edition of the instrument driver can only be installed.
    • When using 64bit OS, the x64 edition of the instrument driver can only be installed, but it contains the x86 (WOW64) edition.

    Download

    Common Library / Documents

    VISA Libraries
    IVI Instrument Driver Programming Guide

    • 2022-07-02 README
      			***********************
      			**** Read Me First ****
      			***********************
      
      Version 1.1.0
      Apr 03, 2017
      
      
      KikusuiPcrle3 IVI Driver for KIKUSUI PCR-LE/LE2 series AC Power Supply
      (This driver requires 3-phase adapters being installed in the PCR-LE series.)
      ---------------------------------------------------------------------------
      
      
      Supported Instruments
      ---------------------
      PCR500LE,PCR1000LE,PCR2000LE,PCR3000LE,PCR4000LE,PCR6000LE,PCR9000LE
      PCR6000LE2,PCR9000LE2,PCR12000LE2,PCR18000LE2,PCR27000LE2
       
      Installation
      -------------
        System Requirements: The driver installation will check for the
        following requirements.  If not found, the installer will either
        abort, warn, or install the required component as appropriate.
      
        Supported Operating Systems:
          Windows 7
          Windows 8
          Windows 10
      
        Shared Components
          Before this driver can be installed, your computer must already
          have the IVI Shared Components installed.
      	(This driver requires the IVI Shared Component since version 2.3.)
      
          The IVI Shared Components installer is available from: 
          http://www.ivifoundation.org/shared_components/Default.aspx
      
      
        VISA-COM
          Any compliant implementation is acceptable. Typically, VISA-COM is
          installed with VISA and other I/O library modules. 
      
      
      Additional Setup
      ----------------
        .NET Framework
          The .NET Framework itself is not required by this driver. If you
          plan to use the driver with .NET, the minimal .NET framework version is 2.0.
      
        The .NET Framework requires an interop assembly for a COM
        server. A Primary Interop Assembly, along with an XML file for
        IntelliSense is installed with the driver. The driver's PIA, along
        with IVI PIAs are installed, by default, in:
        :Program FilesIVI FoundationIVIBinPrimary Interop Assemblies
      
        The PIA is also installed into the Global Assembly Cache (GAC) if
        you have the .NET framework installed.
      
      Help File
      ---------
        The help file (KikusuiPcrle.chm) is located in the directory:
         :Program FilesIVI FoundationIVIDriverskipcrle
      
      Revision History
      ----------------
        Version     Date         Notes
        -------   ------------   -----
          1.0.0   Feb 06, 2012   Initial public release
          1.1.0   Apr 03, 2017   Support PCR-LE2 series
      
      IVI Compliance
      --------------
      IVI-COM/IVI-C Specific Instrument Driver
      IVI Instrument Class: IviACPwr
      
      IviACPwrBase                    yes
      IviACPwrArbWaveform             no
      IviACPwrCurrentProtection       yes
      IviACPwrCurrentRamp             no
      IviACPwrDCGeneration            yes
      IviACPwrExternalSync            yes
      IviACPwrFrequencyRamp           no
      IviACPwrImpedance               yes
      IviACPwrMeasurement             yes
      IviACPwrPhase                   no
      IviACPwrVoltageProtection       yes
      IviACPwrVoltageRamp             no
      
      Optional Features: 
      Interchangeability Checking     no
      State Caching                   no
      Coercion Recording              no
      
      Driver Identification: 
      Vendor:                         Kikusui
      Description:                    PCR-LE/LE2 series AC Power Supply
      Revision:                       1.1.0
      Component Identifier:           KikusuiPcrle3.KikusuiPcrle3
      
      Hardware Information:
      Instrument Manufacturer:        Kikusui
      Supported Instrument Models:    PCR500LE,PCR1000LE,PCR2000LE,PCR3000LE,PCR4000LE,PCR6000LE,PCR9000LE,PCR6000LE2,PCR9000LE2,PCR12000LE2,PCR18000LE2,PCR27000LE2
      Supported Bus Interfaces:       LAN, USB, Serial
      
      32-bit Software Information:
      Supported Operating Systems:    Windows 7/8/10
      Support Software Required:      VISA
      Source Code Availability:       Source code included with driver.
      
      64-bit Software Information:
      Supported Operating Systems: 	Windows 7/8/10
      Support Software Required:      VISA
      Source Code Availability:       Source code included with driver.

    高機能交流安定化電源PCR-LE/LE2シリーズ用ソフトウェア「Avionics Test Software:SD012-PCR-LE」販売開始のおしらせ

    2013年9月30日

    菊水電子工業(本社:神奈川県横浜市 代表取締役社長:小林一夫)は、当社交流電源用規格試験ソフトウェア「Avionics Test Software(アビオニクス試験用ソフトウェア):SD012-PCR-LE」を、 2013年10月より販売開始いたします。

    近年、航空機の電子化が加速しており、それにともない搭載装置・機器の電源環境試験ニーズが高まるものと予想されます。特に航空機には、電源供給系統が複数あり(単相交流・三相交流・直流およびその電圧・周波数の組合わせで7種)、航空機装置試験用電源を設備する課題として、系統別に試験電源を用意するコストとスペースおよび操作オペレーションがあげられます。当社のPCR-LE/LE2シリーズは、単相交流・三相交流はもちろん、直流出力にも対応。そこに規格試験を簡単に実行できるソフトウェアをご用意することで、コストパフォーマンスに優れた電源規格試験ソリューションを構築できるものと考え、当ソフトウェア「Avionics Test Software:SD012-PCR-LE」を開発し、ご提供するはこびとなりました。

    対応する試験規格は、民間用規格「RTCA/DO-160 」、「JIS W0812」および軍事用規格「MIL-STD-704 」です。 当ソフトウェアは、各規格の試験パターンをライブラリとして用意しており、適用したい規格の結線方式・試験項目を選択するだけで容易に試験が実行できるようになっています。また、試験パラメータの編集も可能なため、開発評価などで試験マージンを変えたいといった使い方にも対応できます。

    特長

    • 対応規格:民間用規格「RTCA DO-160F/G」 、民間用規格「JIS W0812:2004」、軍事用規格「MIL-STD-704A/E/F」
    • 航空機器用規格試験が簡単に設定可能
    • 試験パターンのメモリー機能により、効率的な試験設定が可能
    • 試験条件のレポート作成機能により、試験履歴の保存が可能
    • 試験パラメータの編集機能により、開発評価での利用も可能
    • LANによる遠隔コントロールが可能

    ※注:当ソフトウェアの使用には、当社交流電源PCR-LE/LE2シリーズ(本体ファームウェアver.4.2以上)およびパソコン(WindowsXP以上)が必要です。

    販売データ

    ●製品名(形名):
    Avionics Test Software(SD012-PCR-LE)
    ●市場:
    航空機および航空機搭載機器メーカ
    ●出荷開始:
    2013年10月〜
    ●本体標準価格:
    ¥198,000(税込¥207,900)

    本件に対するお問い合わせ

    菊水電子工業株式会社 営業推進部 販売促進課
    〒224-0023 横浜市都筑区東山田1-1-3
    TEL:045-593-7580 FAX:045-593-8672
    E-mail:hanbai@kikusui.co.jp

    keyboard_arrow_leftプレスリリース

    お気軽にご質問・ご相談ください

    製品カタログダウンロード一覧を見る
    導入検討中・見積り依頼メールで問い合わせる
    close

    Search

    Search for articles on this site

    Search for Products.  サポート記事から検索