TOS9300_绝缘电阻测试部-测量功能

项目TOS9300TOS9301TOS9301PDTOS9303TOS9303LC
电压表测量范围负极:0 Vdc ~ -1.2 kVdc,正极:0 Vdc ~ 7.5 kVdc
分辨率0.1 V
精度负极:±(1 % of reading + 0.001 kV),正极:±(1.2 % of reading + 0.001 kV)
电阻表测量范围0.001 MΩ ~ 100.0 GΩ(Max额定电流的1mA到5nA的范围内)
精度 *1 *2 (GND 为 Guard 时) ( i :测量电流) (R :测量电阻)5nA≦i≦50nA*3500.000 mΩ ≦ R < 1.000 GΩ:±(15 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ: ±(15 % of reading + 5 MΩ)
10.000 GΩ ≦ R ≦ 100.000 GΩ:±(20 % of reading + 200MΩ)
50 nA < i ≦ 100 nA*3200.000 MΩ ≦ R < 1.000 GΩ:±(10 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ:±(10 % of reading + 5 MΩ)
10.000 GΩ ≦ R < 50.000 GΩ:±(10 % of reading + 50 MΩ)
50.000 GΩ ≦ R ≦ 100.000 GΩ:±(20 % of reading + 200 MΩ)
100 nA < i ≦ 200 nA*4100.000 MΩ ≦ R < 1.000 GΩ:±(7 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 2.000 GΩ:±(7 % of reading + 5 MΩ)
2.000 GΩ ≦ R < 10.000 GΩ:±(7 % of reading + 10 MΩ)
10.000 GΩ ≦ R < 50.000 GΩ:±(7 % of reading + 100 MΩ)
200 nA < i ≦ 1 µA*410.000 MΩ ≦ R < 100.000 MΩ:±(5 % of reading + 0.05 MΩ)
100.000 MΩ ≦ R < 1.000 GΩ:±(5 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ:±(5 % of reading + 5 MΩ)
10.000 GΩ ≦ R < 25.000 GΩ:±(5 % of reading + 50 MΩ)
1 µA < i ≦ 1 mA*40.001 MΩ ≦ R < 10.000 MΩ:±(2 % of reading + 0.003 MΩ)
10.000 MΩ ≦ R < 100.000 MΩ:±(2 % of reading + 0.03 MΩ)
100.000 MΩ ≦ R < 1.000 GΩ:±(2 % of reading + 0.3 MΩ)
1.000 GΩ ≦ R < 5.000 GΩ:±(2 % of reading + 3 MΩ)
精度 *5 (GND 为 Low 时) ( i :测量电流) (R :测量电阻)5nA≦i≦50nA*3500.000 mΩ ≦ R < 1.000 GΩ:±(25 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ:±(25 % of reading + 5 MΩ)
10.000 GΩ ≦ R ≦ 100.000 GΩ:±(30 % of reading + 200 MΩ)
50 nA < i ≦ 100 nA*3200.000 MΩ ≦ R < 1.000 GΩ:±(20 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ:±(20 % of reading + 5 MΩ)
10.000 GΩ ≦ R < 50.000 GΩ:±(20 % of reading + 50 MΩ)
50.000 GΩ ≦ R ≦ 100.000 GΩ:±(30 % of reading + 200 MΩ)
100 nA < i ≦ 200 nA*4100.000 MΩ ≦ R < 1.000 GΩ:±(10 % of reading + 0.5 MΩ)
100.000 MΩ ≦ R < 2.000 GΩ:±(10 % of reading + 5 MΩ)
2.000 GΩ ≦ R < 10.000 GΩ:±(10 % of reading + 10 MΩ)
10.000 GΩ ≦ R < 50.000 GΩ:±(10 % of reading + 100 MΩ)
200 nA < i ≦ 1 µA*410.000 MΩ ≦ R < 100.000 MΩ:±(5 % of reading + 0.05 MΩ)
100.000 MΩ ≦ R < 1.000 GΩ:±(5 % of reading + 0.5 MΩ)
1.000 GΩ ≦ R < 10.000 GΩ:±(5 % of reading + 5 MΩ)
10.000 GΩ ≦ R < 25.000 GΩ:±(5 % of reading + 50 MΩ)
1 µA < i ≦ 1 mA*40.001 MΩ ≦ R < 10.000 MΩ:±(2 % of reading + 0.003 MΩ)
10.000 MΩ ≦ R < 100.000 MΩ:±(2 % of reading + 0.03 MΩ)
100.000 MΩ ≦ R < 1.000 GΩ:±(2 % of reading + 0.3 MΩ)
1.000 GΩ ≦ R < 5.000 GΩ:±(2 % of reading + 3 MΩ)
保持功能在 PASS判断显示中保持测试结束时的测量电阻值
偏置取消功能抵消输出电缆间等测量不需要的绝缘电阻,Max 2000 GΩ。有 OFF功能。

*1 湿度 70 %rh 以下(无结露)、没有测试导线摇晃等外部干扰时。 *2 在多湿环境下将接地方式(GND)设置为Low时,从本产品内部的高电压配线部及本产品和EUT间的高电压配线部向接地产生泄漏电流。 根据选配的高压扫描器 TOS9320 的使用及配线情况,该泄漏电流有数 nA ~数十 μA,对测量精度有很大影响。启用补偿进行测量,可以降低泄漏电流的影响。 *3 100 V 以下的测量时,精度加上 10 %。 *4 100 V 以下的测量时,精度加上 5 %。 *5 在湿度 50 %rh 以上的环境下,将测量电流限定为 100 nA 以上(无结露)、没有测试导线摇晃等外部干扰、补偿已启用时。